Z-2248-2021: EJM4011 Jamshidi Evolve" Bone Marrow Biopsy/Aspiration Needle with Specimen Cradle. 11G x 10cm

Recall record Z-2248-2021 for Bard Peripheral Vascular Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2248-2021
Company in FDA record
Bard Peripheral Vascular Inc
FDA record creation date
Aug 12, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to complaints received regarding incomplete/open packaging seals potentially causing product to non-sterile.

Source facts

Action
On 07/15/2021, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Notification via FedEx to customers/distributors informing them that the firm has confirmed a complaint regarding an open packaging seal, breaking the sterile barrier. Customers are instructed to: 1. Please check all inventory locations within their institution for the product listed in Table A and immediately destroy all affected product remaining in their possession. The product should have the sterile barrier broken and destroyed per their local facilitys destruction process. 2. Share this notice with any users of the product within their facilities to ensure they are also aware of this Urgent Medical Device Correction. 3. If they purchased this product from a distributor, contact their distributor for further instructions. 4. Complete and return the attached Customer Response Form to the BD contact noted on the form confirming acknowledgement of the notification, whether or not they have any affected product, so that the Recalling Firm may acknowledge their receipt of this notification and process your replacement. 5. Please contact their Recalling Firm's representative to assist in this process or they may also contact the number below for further assistance. The Recalling Firm will: 1. Issue replacements for all customers affected by the recall following receipt of the completed Customer Response Form. 2. taken action to prevent recurrence of this product issue. For questions or assistance contact (Recall Questions/Product Complaints/Technical Questions): North American Regional Complaint Center 1-844-8BD- LIFE (1-844-823-5433), Say Recall when prompted M-F 8am - 5pm CT
Initiated
2021-06-23
Root cause
Packaging process control
Distribution
U.S Nationwide Distribution in the states of .: AZ, CT, DC, FL, IL, MA, MI, MO, NY, OK, TX, and VA O.U.S.: Australia, Belgium, China, EMEA, Japan, Mexico, Saudi Arabia, Singapore, Taiwan
Recall number
Z-2248-2021
Recall status
Open, Classified
CDRH record ID
188373
FDA record created
2021-08-12
Recall event number
88314

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗