Z-2249-2017: Arsenal Spinal Fixation System, Set Screw, Part No. 47027 Product Usage: The Arsenal Spinal Fixation Sy

Recall record Z-2249-2017 for Alphatec Spine, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2249-2017
Company in FDA record
Alphatec Spine, Inc.
FDA record creation date
Jun 6, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Alphatec Spine is recalling the Arsenal Spinal Fixation System Set Screw due to a trend in set screw postoperative disengagement from the screw body.

Source facts

Action
An Urgent Medical Device Recall notification letter dated 5/8/17 will be sent to customers to inform them that as a result of discussions with the FDA regarding a related recall of Arsenal Set Screw part number 47127, Alphatec Spine has made the decision to recall the original Arsenal Set Screw part number 47027. This action will remove all remaining pieces of part number 47027 from the domestic (U.S.) market and the Japan market. Arsenal set screw part number 47127 was recalled in September 2016. Arsenal set screw part number 4 7027 was not included in the recall because the complaints associated with 47027 were within acceptable safety limits and remains so presently. Alphatec has received eight (8) complaints for disengagement of set screw 47027 since the product was introduced to the market on June 14, 2014, which equates to a disengagement failure rate for P/N 47027 of 0.095%. Customers are informed of the actions to be taken. Customers are instructed to fill out the last page of recall letter to confirm that you have read this notification and have taken all necessary actions as described in this notification.
Initiated
2017-05-05
Root cause
Under Investigation by firm
Terminated
2017-06-06
Distribution
Worldwide Distribution - US Nationwide in the states of (TX, AL, UT, FL, IA, SD, NV, OH, CA, MN, CO, AZ, OK, MI, VA, KY, and Puerto Rico) and country of Japan
Recall number
Z-2249-2017
Recall status
Terminated
CDRH record ID
155473
FDA record created
2017-06-06
Recall event number
77220

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗