Open original FDA record ↗Recall record
- FDA identifier
- Z-2250-2012
- Company in FDA record
- Newport Medical Instruments Inc
- FDA classification date
- Apr 14, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class I
Reason stated in the source
May emit a continuous high priority alarm and the ventilator may stop ventilating, due to a component failure on the control board.
Source facts
- Action
- Newport Medical sent an Urgent Field Correction Notice letter dated August 3, 2012 to all affected customers, via DHL and FedEx courier services. The letter identified the affected product, reason for recall and actions to be taken. Customers were instructed to immediately stop using the affected ventilator and retrieve the replacement kit provided free of charge. In addition, the letter states that if the product was further distributed, to contact their customers and notify them of the recall. Customers were asked to complete and return the enclosed Replacement form as replacements are completed. For questions contact Newport Medical Technical Support Department at 800-451-3111, extension 282, for U.S. customers; 714-427-5811, extension 500, for international customers.".
- Initiated
- 2012-08-03
- Root cause
- Nonconforming Material/Component
- Terminated
- 2014-04-14
- Distribution
- Worldwide distribution: USA (nationwide) and countries of: Brazil, Chile, Egypt, France, Hong Kong, India, Japan, Norway, Philippines and South Africa.
- Recall number
- Z-2250-2012
- Recall status
- Terminated
- CDRH record ID
- 111914
- FDA classified
- 2014-04-14
- Recall event number
- 62856