Z-2250-2024: Brand Name: MEGADYNE" MEGA SOFT" Dual Reusable Patient Return Electrode Product Name: MEGADYNE" MEGA SOFT

Recall record Z-2250-2024 for Megadyne Medical Products, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2250-2024
Company in FDA record
Megadyne Medical Products, Inc.
FDA classification date
Jul 3, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

To limit the use of the Mega Soft Pads to patients aged 12 or older to mitigate the potential risk of patient burns identified with use of Mega Soft pads.

Source facts

Action
An URGENT MEDICAL DEVICE CORRECTION letter, dated 6/17/24, was mailed to consignees. Consignees are instructed to share the recall notification with all users of affected devices and confirm their understanding, post a copy of the communication for user awareness, and forward the recall notification to any facility to which product was shipped. The notification reads that Mega Soft and Mega 2000 devices are only to be used for patients 12 years old. The notification also reminds consignees to follow proper cleaning, placement, and setup steps. Consignees are asked to complete and return the provided Business Reply Form by fax to 844-441-2732 or by email to Ethicon8367@sedgwick.com. Customers with any questions about the communication can contact Ethicon at 1-877-384-4266 from Monday through Friday from 8:00 AM EST to 5:00 PM EST. UPDATE: Starting in September 2024, the firm communicated to new and existing consignees reiterating the use of affected devices in patients aged 12 years and older. Additionally, the firm's letter informs consignees that the eIFU of devices indicates that no additional materials should be placed between the patient and Mega Soft pad (example: no sheet).
Initiated
2024-06-17
Root cause
Unknown/Undetermined by firm
Distribution
Domestic: Nationwide distribution International: AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, CANADA, CHINA, GERMANY, GREECE, INDIA, ITALY, NETHERLANDS, NORWAY, JAPAN, SAUDI ARABIA, SINGAPORE, SOUTH KOREA, SPAIN, THAILAND, TURKEY, UNITED KINGDOM, UNITED STATES,
Recall number
Z-2250-2024
Recall status
Open, Classified
CDRH record ID
208457
FDA classified
2024-07-03
Recall event number
94837

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗