Open original FDA record ↗Recall record
- FDA identifier
- Z-2253-2014
- Company in FDA record
- Ortho Development Corporation
- FDA classification date
- Aug 17, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back into the pouch after sealing which increases the risk that the material properties of the implant would be substantially degraded due to oxidation.
Source facts
- Action
- Ortho Development notified all affected consignees/distributors of the recall via email on July 11, 2014 which included a list of the affected product (part numbers, lot numbers and quantities) under their control. An RMA number and return shipping label were also included with the email. Emails were followed up with phone calls to verify receipt of the notification.
- Initiated
- 2014-07-11
- Root cause
- Packaging process control
- Terminated
- 2014-10-06
- Distribution
- Worldwide Distribution - US Nationwide and the country of Japan. No military/govt/VA consignees.
- Recall number
- Z-2253-2014
- Recall status
- Terminated
- CDRH record ID
- 128975
- FDA classified
- 2014-08-17
- Recall event number
- 68898