Open original FDA record ↗Recall record
- FDA identifier
- Z-2253-2020
- Company in FDA record
- Covidien Llc
- FDA record creation date
- Jun 3, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The product is recalled due to a manufacturing assembly error that may cause the staples to fail to properly advance within the staple cartridge when the instrument handle is compressed.
Source facts
- Action
- All consignees were notified by letter titled "Urgent Medical Device Recall" delivered via Federal Express and/or certified mail beginning May 12, 2020. The letter informs customers of the recall of specific item codes and production lots of Multifire hernia staplers due to a manufacturing assembly error. Use of a device with this condition may result in the failure of staples to properly advance within the staple cartridge when the instrument handle is compressed. Customers are instructed to take the following actions: " Immediately quarantine and discontinue use of the affected item code and lot. " Return affected product. " Promptly forward the information from the customer communication to any recipients to whom they have distributed any of the affected product. " Complete the Recalled Product Return Form even if you do not have inventory. If you have any questions about this event, you may contact Medtronic via email at rs.assurancequality@medtronic.com
- Initiated
- 2020-05-12
- Root cause
- Process control
- Distribution
- Worldwide distribution.
- Recall number
- Z-2253-2020
- Recall status
- Open, Classified
- CDRH record ID
- 181547
- FDA record created
- 2020-06-03
- Recall event number
- 85689