Simpleware Product Group, SYNOPSYS NORTHERN EUROPE
FDA record creation date
Aug 12, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II
Reason stated in the source
A issue (bug) has been identified with the interface and image software which could result in anatomical orientation tags/labels being displayed incorrectly. This could result in operator misunderstanding and errors during resampling or re-alignment of image data
Source facts
Action
Synopsys sent a "Urgent Field Safety Notice" on June 18, 2021 via E-Mail. In addition to informing consignees about the recall, the firm request consignees take the following actions:
1. The user should stop using Simpleware ScanIP Medical R-2021.03 and earlier versions and immediately upgrade to the latest version (S-2021.06a/S-2021.06) of the software on SolvNetPlus.
2. The user should also uninstall and/or delete previous installers.
3. If this is not immediately possible, the user should always remove all DICOM tags before:
" Using the Register datasets tool through the GUI or scripting.
" Or applying any image rotation transforms (e.g. align or flip) through the GUI or scripting. This action can be completed by using the DICOM tags tool in the General section of the Home tab.
4. Complete the customer reply form by July 2, 2021 and submit to via
Email simpleware-qms@synopsys.com
Customer Helpline +44(0)1392 428752
Postal Address: Simpleware Product Group, Synopsys Northern Europe Ltd., Bradninch Hall, Castle Street, Exeter, Devon, EX4 3PL, UK
Web Portal: https://www.synopsys.com/simpleware.html
Fax: N/A
Initiated
2021-06-04
Root cause
Software Manufacturing/Software Deployment
Terminated
2022-09-29
Distribution
US: Nationwide Distribution in the states of CA, FL
and countries of : France, United Kingdom, Germany, Australia
Recall number
Z-2253-2021
Recall status
Terminated
CDRH record ID
188131
FDA record created
2021-08-12
Recall event number
88207
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.