Skip to content

Public FDA record

Z-2254-2021: Simpleware ScanIP medical software. For transfer of imaging information from a medical scanner to an out

Recall record Z-2254-2021 for Simpleware Product Group, SYNOPSYS NORTHERN EUROPE. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2254-2021
Company in FDA record
Simpleware Product Group, SYNOPSYS NORTHERN EUROPE
FDA record creation date
Aug 12, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A issue (bug) has been identified with the interface and image software which could result in anatomical orientation tags/labels being displayed incorrectly. This could result in operator misunderstanding and errors during resampling or re-alignment of image data

Source facts

Action
Synopsys sent a "Urgent Field Safety Notice" on June 18, 2021 via E-Mail. In addition to informing consignees about the recall, the firm request consignees take the following actions: 1. The user should stop using Simpleware ScanIP Medical R-2021.03 and earlier versions and immediately upgrade to the latest version (S-2021.06a/S-2021.06) of the software on SolvNetPlus. 2. The user should also uninstall and/or delete previous installers. 3. If this is not immediately possible, the user should always remove all DICOM tags before: " Using the Register datasets tool through the GUI or scripting. " Or applying any image rotation transforms (e.g. align or flip) through the GUI or scripting. This action can be completed by using the DICOM tags tool in the General section of the Home tab. 4. Complete the customer reply form by July 2, 2021 and submit to via Email simpleware-qms@synopsys.com Customer Helpline +44(0)1392 428752 Postal Address: Simpleware Product Group, Synopsys Northern Europe Ltd., Bradninch Hall, Castle Street, Exeter, Devon, EX4 3PL, UK Web Portal: https://www.synopsys.com/simpleware.html Fax: N/A
Initiated
2021-06-04
Root cause
Software Manufacturing/Software Deployment
Terminated
2022-09-29
Distribution
US: Nationwide Distribution in the states of CA, FL and countries of : France, United Kingdom, Germany, Australia
Recall number
Z-2254-2021
Recall status
Terminated
CDRH record ID
188132
FDA record created
2021-08-12
Recall event number
88207

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗