Z-2255-2023: NeuMoDx Cartridge-used for extraction, purification, amplification and detection of nucleic acids on the

Recall record Z-2255-2023 for NeuMoDx Molecular Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2255-2023
Company in FDA record
NeuMoDx Molecular Inc
FDA classification date
Jul 28, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Top label on some of the NeuMoDx cartridges within these lots may have been incorrectly applied during the manufacturing process, resulting in encroachment into the PCR region. The misplaced label may interfere with the module s optics performance, thereby increasing the risk for false positive results

Source facts

Action
Qiagen issued customers "Urgent: Medical Device Correction" Letter via email on 5/24/23. Letter states reason for recall, health risk and action to take: If you have remaining stock of cartridges LOT 117514, 117524 or 117528, REF 100100, do not use it. Please contact QIAGEN Technical Service for a free-of-charge replacement. Dispose of the product LOT 117514, 117254, and 117528 in accordance with your national and local safety and environmental regulations. If you have already used NeuMoDx Cartridges from this LOT, please review all results obtained with the laboratory director and assess whether retesting is required. Review this notice with your laboratory/medical director. Important: Forward this information to all individuals and departments within your organization using the above listed cartridges. If you are not the end user, please forward this notice to the product end user. Complete Acknowledgement of Receipt attached to this letter by as soon as possible. If you have any questions or concerns, please contact your local QIAGEN Technical Service Department through any of the following: www.qiagen.com/QIAGEN-Subsidiaries.
Initiated
2023-05-24
Root cause
Employee error
Distribution
US Nationwide Distribution to states of: Kansas, Florida, Texas, Ohio, New York, Michigan, Pennsylvania; and OUS (Foreign) to countries of: AT, BE, CH, DE, ES, FI, FR, GB, IT, LV, PL, RO, SE
Recall number
Z-2255-2023
Recall status
Open, Classified
CDRH record ID
201821
FDA classified
2023-07-28
Recall event number
92615

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗