Z-2256-2018: C2 CryoBalloon Focal Catheter, REF FG-1009. C2 CryoBalloon Ablation System is used as a cryosurgical too

Recall record Z-2256-2018 for PENTAX of America Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2256-2018
Company in FDA record
PENTAX of America Inc
FDA classification date
Jun 5, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The catheter alignment arrow printed onto the strain relief is not aligned with the detachment feature of the catheter connector, resulting in the potential for the catheter to be attached to the controller in the detach orientation. In this orientation, the catheter is not securely attached to the controller and may be partially (~1cm) ejected from controller.

Source facts

Action
On 3/30/2018, C2 Therapeutics issued URGENT FIELD CORRECTION notices to customers via certified mail, return receipt requested. Customers are advised to take the following Action: - Forward a copy of the notice and the enclosures to the department in which is referenced in the notice. - Inspect stock and quarantine products identified within the notice. - The end user of the affected products should complete and return the response form via email to RMA@c2therapeutics.com. The response form contains an RMA number to allow for the return of devices affected by this field action, and enclosed in the package is a Federal Express shipping label for the returned merchandise. Customers with questions can call 650-318-5899, Monday through Friday 8:30 AM - 5 PM, PST.
Initiated
2018-03-30
Root cause
Vendor change control
Terminated
2021-04-28
Distribution
Distribution US nationwide.
Recall number
Z-2256-2018
Recall status
Terminated
CDRH record ID
164786
FDA classified
2018-06-05
Recall event number
80180

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗