Z-2256-2020: ProxiDiagnost N90, UDI 706100

Recall record Z-2256-2020 for Philips North America, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2256-2020
Company in FDA record
Philips North America, LLC
FDA record creation date
Jun 3, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The thermo switches in the three-phase transformer, which normally aid in powering down the System Power Distribution Unit, may be incorrectly installed and not working. If the transformer overheats due to a first failure and the thermo switch is not activated, the device may begin to generate smoke.

Source facts

Action
On April 23, 2020, the firm distributed Urgent Field Safety Notice letters to affected customers. Customers were informed of the product issue. According to the notice, the customer can use the system according to the IfU (Instruction for Use). The user should pay attention to whether it smells of smoke in the room. If this is the case, immediately shut down the system according to IFU or (if present) use emergency off switch in room, remove patient, stop using the system for further examinations and contact customer service. Philips plans to perform an On-Site check of the SPDU thermo contact at affected systems, which will eliminate these issues. Should you need to communicate with Philips with regard to this program, please reference Field Change Order 70600101. If you need any further information or support concerning this issue, please contact your local Philips representative.
Initiated
2020-04-23
Root cause
Process control
Terminated
2023-03-30
Distribution
Domestic distribution nationwide. Foreign distribution to Chile and Germany.
Recall number
Z-2256-2020
Recall status
Terminated
CDRH record ID
181316
FDA record created
2020-06-03
Recall event number
85603

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗