Z-2256-2023: The following Pro.specta Emission Computed Tomography Systems that were manufactured with an SLD Look Ahe

Recall record Z-2256-2023 for Siemens Medical Solutions USA, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2256-2023
Company in FDA record
Siemens Medical Solutions USA, Inc.
FDA classification date
Jul 28, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During a whole-body planar scan with auto-contour, there is a possibility the short-linear drive (SLD) look-ahead sensors may trigger earlier than expected and may result in detector 1 positioning further from the patient than needed leading to a reduction in image resolution.

Source facts

Action
The firm issued undated letters flagged Customer Safety Advisory Notice CAN 001-2023 on 7/12/2023 via email, certified mail, or, if necessary, hand delivery. The letter explains the issue, when the malfunction occurs, the potential risks, and how the consignee can help to avoid the potential risk of the issue. The letter informs the consignee that Siemens Healthineers is working to address the issue through a scheduled service visit and the consignee's local service organization will begin contacting the consignees in the 4th quarter of 2023 to schedule this work. In the meantime, the notice is to be placed in the Symbia Pro.specta Operator Manual and disseminated to all operators of the scanner. No response form was enclosed. The recalling firm uses an SAP system for tracking purposes. For the letters that are delivered via email, the consignee verifies receipt through an adobe signature. For the letters delivered via certified mail, there is a return receipt.
Initiated
2023-07-12
Root cause
Device Design
Distribution
Distribution was made to ID, KS, MA, NH, OH, TX, and VA. There was no government/military distribution. Foreign distribution was made to Argentina, Australia, Austria, Belgium, Canada, Colombia, Denmark, Finland, France, Germany, Hong Kong, Israel, Italy, Japan, Netherlands, Portugal, Singapore, Switzerland, Turkey, United Arab Emirates, and United Kingdom.
Recall number
Z-2256-2023
Recall status
Open, Classified
CDRH record ID
202218
FDA classified
2023-07-28
Recall event number
92704

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗