Open original FDA record ↗Recall record
- FDA identifier
- Z-2257-2020
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- Jun 3, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Using Quantification Application (QVA/QCA) on DSA images may lead to a failure in vessel detection or incorrect quantification of vessel detection and result in an incorrect diagnosis and inappropriate treatment of the patient
Source facts
- Action
- Siemens issued Urgent Medical Device Correction (AX035/20/S) dated 4/23/20. The letter states reason for recall health risk and action to take: We urgently recommend not to use Quantification Application (QVA/QCA) on DSA images until the software problem has been resolved. If Quantification Application has already been used on DSA images in the past, please verify the results and diagnostic evaluation if applicable. Siemens will correct the error with a software patch via Update Instruction AX034/20/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436.
- Initiated
- 2020-04-23
- Root cause
- Software design
- Terminated
- 2020-10-26
- Distribution
- AZ, IL, UT, WI
- Recall number
- Z-2257-2020
- Recall status
- Terminated
- CDRH record ID
- 181349
- FDA record created
- 2020-06-03
- Recall event number
- 85615