Skip to content

Public FDA record

Z-2257-2020: ARTIS icono systems with Quantification Application SW (QVA/QCA) on a standalone XWP computer, an interve

Recall record Z-2257-2020 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2257-2020
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
Jun 3, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Using Quantification Application (QVA/QCA) on DSA images may lead to a failure in vessel detection or incorrect quantification of vessel detection and result in an incorrect diagnosis and inappropriate treatment of the patient

Source facts

Action
Siemens issued Urgent Medical Device Correction (AX035/20/S) dated 4/23/20. The letter states reason for recall health risk and action to take: We urgently recommend not to use Quantification Application (QVA/QCA) on DSA images until the software problem has been resolved. If Quantification Application has already been used on DSA images in the past, please verify the results and diagnostic evaluation if applicable. Siemens will correct the error with a software patch via Update Instruction AX034/20/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436.
Initiated
2020-04-23
Root cause
Software design
Terminated
2020-10-26
Distribution
AZ, IL, UT, WI
Recall number
Z-2257-2020
Recall status
Terminated
CDRH record ID
181349
FDA record created
2020-06-03
Recall event number
85615

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗