Z-2257-2023: MEGADYNE Suction Coagulators, Product Code 004125 (Suction Coagulator, Handswitching, 10 Fr, 6 inch (15.2

Recall record Z-2257-2023 for Megadyne Medical Products, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2257-2023
Company in FDA record
Megadyne Medical Products, Inc.
FDA classification date
Jul 28, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An issue with the Suction Coagulator product family may lead to fluid ingress into the handpiece. Fluid ingress can lead to intermittent device activation, non-activation, or self-activation when plugged into the electrosurgery unit (ESU).

Source facts

Action
An "URGENT: MEDICAL DEVICE RECALL (REMOVAL)" letter dated 6/15/23 was sent to customers. If you have additional questions regarding this recall or to report any customer complaints, please contact the Ethicon Resource Department at 1-877-ETHICON (1-877-384-4266). The Ethicon Resource Department is open Monday through Friday, 8:00 AM to 5:00 PM ET. ACTION REQUIRED 1. Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s). If you have product subject to this recall, please maintain a copy of this notice with the quarantined product and keep a copy for your records. 2. Communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall letter when communicating. 3. Complete the Business Reply Form (BRF) (Attachment 2) confirming receipt of this notice and fax or email to Sedgwick at 866-918-8739 or Ethicon8055@sedgwick.com within three (3) business days. Please return the BRF even if you do not have product subject to this recall. 4. Customers are required to return unused MEGADYNE" Suction Coagulators subject to this recall that are in inventory immediately. To receive credit reimbursement, customers must return product subject to this recall no later than September 30, 2023 to Sedgwick at 866-918-8739 or Ethicon8055@sedgwick.com. Any non-affected product and any product returned after the date specified will not receive credit reimbursement. 5. To return product subject to this recall, photocopy the completed BRF, place it in the box with the product, and affix the pre-paid authorized shipping label included with the recall notification letter. Megadyne will pay for the shipping charges only if the authorized label is used. Extra shipping labels may be ob
Initiated
2023-06-15
Root cause
Under Investigation by firm
Distribution
Worldwide Distributions: US (Nationwide) and OUS (Foreign) in countries of: AUSTRALIA, BELGIUM, CANADA, CHILE, DENMARK, EGYPT, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, NEW ZEALAND, NORWAY, RUSSIA, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, UNITED ARAB EMIRATES, UNITED KINGDOM, VIETNAM
Recall number
Z-2257-2023
Recall status
Open, Classified
CDRH record ID
200953
FDA classified
2023-07-28
Recall event number
92456

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗