Open original FDA record ↗Recall record
- FDA identifier
- Z-2258-2014
- Company in FDA record
- Smith & Nephew, Inc.
- FDA classification date
- Aug 17, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There was an error in manufacturing which resulted in the proximal lag screw hole being positioned incorrectly in several batches of TRIGEN(R); INTERTAN(TM); NAILS.
Source facts
- Action
- Smith & Nephew Issued an Urgent Product Recall dated July 2, 2014 to all affected customers via E-mail and/or FedEx . The notification identified the effective product, problem and actions to be taken. Customers were instructed to: Inspect inventory and locate any unused affected devices, cease distribution and use of those products, quarantine them immediately, complete the last two columns in the Inventory Return Certification Form, and return any affected product to the address provided. For questions call 901-396-2121
- Initiated
- 2014-07-02
- Root cause
- Under Investigation by firm
- Terminated
- 2019-08-20
- Distribution
- Worldwide Distribution - US Nationwide and the countries of AUSTRALIA, CANADA, CHINA, INDIA, UAE, JAPAN, SOUTH AFRICA, SOUTH KOREA, UNITED KINGDOM. AUSTRIA, NORWAY, POLAND, PORTUGAL
- Recall number
- Z-2258-2014
- Recall status
- Terminated
- CDRH record ID
- 128687
- FDA classified
- 2014-08-17
- Recall event number
- 68792