Open original FDA record ↗Recall record
- FDA identifier
- Z-2258-2020
- Company in FDA record
- GE Healthcare, LLC
- FDA classification date
- Jun 3, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
GE Healthcare has identified that a small number of Nuclear Medicine systems are operating without currently updated software. In a recent event, a patient s finger was injured during the Unload process while the automatic Unload feature was being used; the software on the system was not current.
Source facts
- Action
- GE Healthcare identified that a small number of Nuclear Medicine systems are operating without currently updated software.
- Initiated
- 2018-11-30
- Root cause
- Other
- Distribution
- Domestic Distribution: AL, AR, AZ, CA, CO, CT, DC, DE, FL,GA, IA, IL, IN, KY, LA,MA, MD, ME, MI, MN, MO, MS, NC, NE, NM, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT,VA, WI and WV. Foreign Distribution: Argentina, Australia, Belgium, Brazil, Canada, Colombia, Denmark, France, Germany, Ghana, India, Israel, Italy, Japan, Korea, Malaisya, Mexico, Netherlands, New Zealand, Norway, Poland, Portugal, Singapore, Sweden, Taiwan, Thailand, Turkey and United Kingdom.
- Recall number
- Z-2258-2020
- Recall status
- Open, Classified
- CDRH record ID
- 181672
- FDA classified
- 2020-06-03
- Recall event number
- 85746