Z-2260-2018: The BD Vacutainer Plus Plastic Citrate Blood Collection Tube is an evacuated blood collection tube that p

Recall record Z-2260-2018 for Becton Dickinson & Company. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2260-2018
Company in FDA record
Becton Dickinson & Company
FDA record creation date
Jun 20, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Approximately 3.4% of BD Vacutainer Plus Plastic Citrate Blood Collection Tubes (2.7mL) (light blue top tube) will exhibit over fill draw volumes from ~11-14%.

Source facts

Action
BD began notifying distributors on April 11, 2018. Customer communications were mailed on April 12-13th, 2018. US customers will be sent a letter via UPS beginning April 12, 2018. Distributors were e-mailed on April 11, 2018 and requested that they provide a list of their consignees to facilitate mailing to end users BD Regions will be contacted via email regarding distribution outside of the US Distributors were instructed to return customer lists to BD. Letters will be mailed with UPS tracking, which provides for a 100% effectiveness check. Customer response form will be tabulated. 100% of non-responders will be contacted by phone.
Initiated
2018-04-11
Root cause
Nonconforming Material/Component
Terminated
2019-05-10
Distribution
US/PR-All 50 states and PR Global-AR, BO, BR, BS, BZ, CA, CO, CR, CW, DO, EC, GT, GY, IN, JM, JP, KR, MX, NI, PA, PE, SG, SR, SV, TT, TW, UY, VI Dates of distribution: June 2, 2017 to the present.
Recall number
Z-2260-2018
Recall status
Terminated
CDRH record ID
163870
FDA record created
2018-06-20
Recall event number
79901

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗