Z-2260-2020: da Vinci SP surgical system

Recall record Z-2260-2020 for Intuitive Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2260-2020
Company in FDA record
Intuitive Surgical, Inc.
FDA classification date
Jun 3, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Intuitive has become aware that the da Vinci SP system may trigger a mechanical vibration of the instrument tips and endoscope due to a software anomaly. This issue has the potential to occur only under the following, extremely rare specific conditions: 1) the user is activating Adjust Mode, AND 2) the instrument Arm is near its vertical position limit, AND 3) the user is applying sustained force against the hand control haptic feedback.

Source facts

Action
The firm sent a letter to consignees on April 1, 2020 stating that an Intuitive representative will schedule a site visit to provide a software update as soon as possible. Firm representatives will be available by phone to answer any questions related to this Medical Device Correction. Actions taken by the customer are to: Inform all personnel who are involved with the affected device at your site of the following instructions: 1.If vibrations are sensed at the instrument and/or endoscope tips in Adjust Mode and the instrument arm is near its vertical limit: a.Cease applying sustained force against the hand control haptic feedback, b.Exit Adjust Mode by releasing the Endoscope Control pedal c.Exit following mode by removing your head from the Surgeon Console, or d.Press the Emergency Stop button. Any of these actions will stop the vibration. 2.If mechanical vibration persists, discontinue use of the system and contact Technical Support. 3.Inform affected personnel when the software update has been completed by the Intuitive representative. 4.Complete the attached Acknowledgment Form and return it via fax or email to Intuitive as instructed on the form. 5.Please retain a copy of this letter and the acknowledgment form for your files. Customers with additional questions: North and South America: (800) 876-1310, Option 3 (4 AM to 5 PM PST); South Korea: + 82-02-3271-3200, Opt. 3 (9 AM to 6 PM KSTJ)
Initiated
2019-04-01
Root cause
Software design
Terminated
2021-12-16
Distribution
U.S.: OH, PN, NY, MN, TN, TX, OR, IL, MD, NJ, AL. O.U.S.: South Korea
Recall number
Z-2260-2020
Recall status
Terminated
CDRH record ID
180439
FDA classified
2020-06-03
Recall event number
85293

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗