Skip to content

Public FDA record

Z-2261-2015: Ottobock Pheon 3R62 Knee Joint The 3R62 Pheon is a polycentric prosthetic knee joint intended to be used

Recall record Z-2261-2015 for Otto Bock Healthcare GmbH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2261-2015
Company in FDA record
Otto Bock Healthcare GmbH
FDA classification date
Jul 24, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Otto Bock Healthcare GmbH is voluntarily recalling all 3R62 Pheon polycentric prosthetic knee joints distributed since February 2014 because the locking function may fail under certain conditions which may result in a fall.

Source facts

Action
The firm, Ottobock, sent an "URGENT MEDICAL DEVICE RECALL 3R62 Pheon" letter dated 16 July 2015 to its customers. The letter described the product, problem, and actions to be taken. The customers instructed to read this notice in its entirety; examine your stock for 3R62 Pheon devices; quarantine 3R62 Pheon devices and prevent further distribution; contact Ottobock customer service at (800) 328-4058-option 1 or (512) 806-2000 -Option 1, to obtain a return authorization for those devices, and complete and return the MEDICAL DEVICE RECALL RETURN RESPONSE Acknowledgement and Receipt Form via email to: USRecalls@ottobock.com "Pheon Recall"; fax (801) 974-6718-"Pheon Recall" or mail to: Otto Bock HealthCare LP, ATTN: PHEON RECALL, 3820 W. Great Lakes Drive, Salt Lake City, UT 84120-B7 is re-designing the locking function to prevent this failure from occurring. The re-designed devices will be available for replacement sometime within the next few months. If you have any questions contact Director, Regulatory & Quality Compliance at (801) 974-6618 or email: Lex.Pearce@ottobock.com.
Initiated
2015-07-16
Root cause
Component design/selection
Terminated
2017-06-21
Distribution
Worldwide Distribution: US (nationwide) and countries of: AU, BE, CN, CZ, IT, NL, and SK.
Recall number
Z-2261-2015
Recall status
Terminated
CDRH record ID
138816
FDA classified
2015-07-24
Recall event number
71715

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗