Open original FDA record ↗Recall record
- FDA identifier
- Z-2261-2015
- Company in FDA record
- Otto Bock Healthcare GmbH
- FDA classification date
- Jul 24, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Otto Bock Healthcare GmbH is voluntarily recalling all 3R62 Pheon polycentric prosthetic knee joints distributed since February 2014 because the locking function may fail under certain conditions which may result in a fall.
Source facts
- Action
- The firm, Ottobock, sent an "URGENT MEDICAL DEVICE RECALL 3R62 Pheon" letter dated 16 July 2015 to its customers. The letter described the product, problem, and actions to be taken. The customers instructed to read this notice in its entirety; examine your stock for 3R62 Pheon devices; quarantine 3R62 Pheon devices and prevent further distribution; contact Ottobock customer service at (800) 328-4058-option 1 or (512) 806-2000 -Option 1, to obtain a return authorization for those devices, and complete and return the MEDICAL DEVICE RECALL RETURN RESPONSE Acknowledgement and Receipt Form via email to: USRecalls@ottobock.com "Pheon Recall"; fax (801) 974-6718-"Pheon Recall" or mail to: Otto Bock HealthCare LP, ATTN: PHEON RECALL, 3820 W. Great Lakes Drive, Salt Lake City, UT 84120-B7 is re-designing the locking function to prevent this failure from occurring. The re-designed devices will be available for replacement sometime within the next few months. If you have any questions contact Director, Regulatory & Quality Compliance at (801) 974-6618 or email: Lex.Pearce@ottobock.com.
- Initiated
- 2015-07-16
- Root cause
- Component design/selection
- Terminated
- 2017-06-21
- Distribution
- Worldwide Distribution: US (nationwide) and countries of: AU, BE, CN, CZ, IT, NL, and SK.
- Recall number
- Z-2261-2015
- Recall status
- Terminated
- CDRH record ID
- 138816
- FDA classified
- 2015-07-24
- Recall event number
- 71715