- Recall classification
- Class II
Reason stated in the source
An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications.
Source facts
- Action
- On 6/23/2016, the recalling firm sent written recall notices to their distributors and end users notifying them of the recall. Customers and consignees were instructed to not use the devices, locate and quarantine all affected devices, and to complete the response form.
- Initiated
- 2016-06-23
- Root cause
- Device Design
- Terminated
- 2023-04-11
- Distribution
- Product distributed to medical facilities in the following US states: AL, AZ, CA, CO, CT, FL, IA, IL, KS, LA, MA, MD, MN, NC, OH, PA, PR, TN, TX, UT, and WA. Product was also distributed internationally to the following countries: Australia, Canada, Dominican Republic, Malaysia, Turkey, and United Kingdom
- Recall number
- Z-2261-2016
- Recall status
- Terminated
- CDRH record ID
- 147603
- FDA record created
- 2016-07-22
- Recall event number
- 74043