Skip to content

Public FDA record

Z-2261-2020: Xhibit Central, Model No. 96102 - Product Usage: intended use is to provide clinicians with central monit

Recall record Z-2261-2020 for Spacelabs Healthcare, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2261-2020
Company in FDA record
Spacelabs Healthcare, Inc.
FDA record creation date
Jun 4, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm received reports of loss of audio alarm after a power failure or cable disconnection. The loss of audio for a telemetry patient could delay the recognition of an alarm condition.

Source facts

Action
Urgent Medical Device Correction letters were sent to U.S. customers on August 30, 2019 via priority service, and to foreign customers via email on September 6, 2019. The letter informed customers of the conditions under which alarm audio may be lost, and actions the customer /user can take to minimize risk to patients or users. Customers were requested to weigh the benefits versus the risks when deciding whether or not to continue to use Xhibit Central with software versions 1.3.1, 1.3.2, 1.3.3, and 1.3.4. Customers experiencing a loss of audio were informed that a complete reboot of the Xhibit Central will return the device to normal operation. Customers were requested to circulate this notification to all persons affected by the information it contains. Customers were also informed that the manufacturer would contact them to schedule a time for Spacelabs to update their Xhibit Central software. Customers with questions, requiring information, or technical assistance may contact the Technical Support group at Spacelabs Healthcare, Inc. at 1-800-522-7025 and select 2 for Technical Support.
Initiated
2019-08-30
Root cause
Software Manufacturing/Software Deployment
Terminated
2024-04-18
Distribution
Worldwide distribution - US Nationwide distribution including in the 40 states and Puerto Rico, and multiple countries including Canada.
Recall number
Z-2261-2020
Recall status
Terminated
CDRH record ID
180525
FDA record created
2020-06-04
Recall event number
85325

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗