The markings on the distal face of the instruments are mis-oriented by 180. This includes Anterior and Posterior, which the surgeon uses to correctly place the cut block on the resectioned face of the femur. The cut block is not symmetric, so correct orientation of the instrument in the A/P direction is critical to making proper anterior, posterior, and chamfer cuts to fit the femoral implant.
Source facts
Action
DJO Surgical sent an " Urgent Field Safety Notice" dated June 30, 2015, to the affected customer. The recalling firm requested the customer return the affected product for a replacement. For further questions, please call (512) 834-6302.
Initiated
2015-06-24
Root cause
Component design/selection
Terminated
2016-06-24
Distribution
US Distribution to TX and GA.
Recall number
Z-2262-2015
Recall status
Terminated
CDRH record ID
138622
FDA classified
2015-07-27
Recall event number
71663