Skip to content

Public FDA record

Z-2262-2018: Nitinol Staple, Elastic Staple Large-Asymmetrical, Sterile R, Model #: a) ES-18x15x17; b) ES-18x17x19

Recall record Z-2262-2018 for TriMed Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2262-2018
Company in FDA record
TriMed Inc.
FDA record creation date
Jun 20, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.

Source facts

Action
On 04/13/18, customer recall communication forms were sent via mail informing customers to quarantine effected devices, call the manufacturer to arrange device return at 800-633-7221, and report adverse events to the FDA's MedWatch.
Initiated
2018-04-13
Root cause
Process design
Terminated
2021-01-22
Distribution
U.S.: IL, UT, PA, NJ, WA, TX, NV, CA, NC, MS, KS, MO, MN, FL; OUS: Australia and Sweden
Recall number
Z-2262-2018
Recall status
Terminated
CDRH record ID
164555
FDA record created
2018-06-20
Recall event number
80099

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗