Z-2262-2020: Model Number: 06917259001 UDI: 04015630976010 - Product Usage: The System s intended use environment

Recall record Z-2262-2020 for Ventana Medical Systems Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2262-2020
Company in FDA record
Ventana Medical Systems Inc
FDA record creation date
Jun 5, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm became aware of a cleaning solution leaking issue from the middle staining module of the Ventana HE 600 instrument, used for histologic section samples in anatomic pathology laboratory settings, which caused fluid to travel down to the electrical wiring connector of the Coverslipper solvent shield circuit board. This can cause an electrical short and result in burned components.

Source facts

Action
All US consignee accounts who are utilizing the VENTANA HE 600 system were to receive a copy of the Urgent Medical Device Correction (UMDC) notification, VENTANA HE 600 Instrument - Fluid Leak and Risk of Burned Components, via UPS Ground on 10/16/19. It notified user that Roche Tissue Diagnostics received notification of an issue on the VENTANA HE 600 instrument. A leak from the middle staining module caused fluid to travel down to the electrical wiring connector of the Coverslipper solvent shield circuit board. This caused an electrical short and resulted in burned components. It stated clinical significance that there is a remote possibility that use of the VENTANA HE 600 system would result in adverse health consequences. It advised users that Roche Service Representatives will remove the cable that conducts electricity to the Coverslipper Module LED PCBA, proactively replace the Stainer Exchange Valves, and proactively replace the 65 micron filters in the Cleaning Solution fluidics line. Site visits will be scheduled as appropriate. Customers can continue using the instrument as normal. However, if droplets are observed on the instrument Coverslip Module during normal operation: stop instrument operation, unplug the instrument, and contact your Roche Service Representative immediately. Actions Required: " If customers observe droplets on the instrument Coverslip Module during normal operation: stop instrument operation, unplug the instrument, and contact your Roche Service Representative immediately. " Complete attached fax form and fax or email it according to the instructions on the form. " File the Urgent Medical Device Correction (UMDC) for future reference. Customers can contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-227-2155 if they have questions about the information contained in the UMDC. Adverse events or quality problems experienced with the use of this product may also be reported to
Initiated
2019-10-04
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, MD, MI, MA, MN, MO, MS, MT, NC, NE, NJ, NY, OH, OK, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, China, Columbia, Costa Rica, Croatia, Denmark, Ecuador, El Salvador, Finland, France, Germany, Hong Kong, Ireland, Italy, Japan, Kuwait, Lebanon, Malaysia, Norway, Peru, Poland, Saudi Arabia, Slovakia, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Vietnam.
Recall number
Z-2262-2020
Recall status
Completed
CDRH record ID
181168
FDA record created
2020-06-05
Recall event number
85526

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗