Z-2263-2024: Zyno Medical Z800, Z800F, Z800W, and Z800WF Infusion Pumps

Recall record Z-2263-2024 for Zyno Medical LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2263-2024
Company in FDA record
Zyno Medical LLC
FDA classification date
Jul 3, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Zyno Medical has found that certain Z800, Z800F, Z800W, and Z800WF pumps have faulty power supply modules that will not provide a charge to the rechargeable battery within the pump. This may lead to the battery having insufficient or no charge available to the pump when it is unplugged from the AC mains power.

Source facts

Action
An URGENT: MEDICAL DEVICE CORRECTION notification letter dated 6/4/24 was sent to customers. Recommended Actions When using a system that has been identified as having the power supply module issue, keep them plugged into an external AC mains power source. 1. Actions to Address the Issue During Use Operate the unit on AC mains power source. If the affected system is disconnected from the external AC mains power source and the system presents an audible low battery alarm, re-connect the system to the external AC mains power source. If the affected system is disconnected from the external AC mains power source and immediately shuts down, re-connect to the external AC mains power source, and re-start the infusion therapy. If the system is turned on and it presents an audible low battery alarm prior to initiation of infusion therapy, proceed to a different system for the planned infusion therapy. Note that only the devices listed in this letter are affected. No other products are involved in this action. 2. Actions to Correct the Issue Complete the attached Zyno Medical Z800, Z800F, Z800W, and Z800WF Infusion Pump Power Supply Module Issue Response Verification Form. This will include checking your inventory for the affected products, verifying serial numbers, confirming product location, reporting any transferred product, and signing and dating the completed form. A representative from Zyno Medical s partner company, IntuVie LLC, will be in contact with you to coordinate instructions on exchanging devices to remediate this action. Pass this notice to the appropriate personnel who need to be informed if the identified devices have been transferred to another location. If you have any questions or would like assistance, please contact your local Business Development Manager. OTHER INFORMATION: " feedback@intuvie.com
Initiated
2024-06-04
Root cause
Process control
Distribution
US Nationwide distribution.
Recall number
Z-2263-2024
Recall status
Open, Classified
CDRH record ID
208221
FDA classified
2024-07-03
Recall event number
94768

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗