Skip to content

Public FDA record

Z-2265-2012: Canon Solid State X-ray Imager (Flat Panel/Digital Imager), Digital Radiography, Models CXDI-55C, CXDI-55

Recall record Z-2265-2012 for Canon Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2265-2012
Company in FDA record
Canon Inc.
FDA classification date
Aug 24, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

Screws holding the 55REF circuit board are loosening and roaming inside of the sealed sensor.

Source facts

Action
Canon Inc. notified sales companies including Canon USA on 9/15/2010 with Service Bulletin 10-017, and on 8/25/2011 with Service Bulletin 11-011, and on 1/20/2011 with Service Bulletin 11-001. Sales representatives were to visit customers and check and replace the screws.
Initiated
2010-11-01
Root cause
Nonconforming Material/Component
Terminated
2014-01-07
Distribution
Nationwide Distribution (USA)
Recall number
Z-2265-2012
Recall status
Terminated
CDRH record ID
111794
FDA classified
2012-08-24
Recall event number
62828

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗