A screw set was returned from the field for destruction after being exposed to flood water, which was erroneously reprocessed and returned to the field. Initial investigative testing indicated the components within the set may be cytotoxic.
Source facts
Action
An Urgent Field Safety Notification, dated July 21, 2014, was sent to surgeons explaining the product, problem, and action to be taken with regard to routine post-operative follow-up on patients exposed to the affected product. Questions or concerns should be directed to Leonie Rynn at 610-719-5408 or lrynn1@its.jnj.com.
Initiated
2014-06-30
Root cause
Employee error
Terminated
2015-12-18
Distribution
Distributed in the states of WI and MI.
Recall number
Z-2265-2014
Recall status
Terminated
CDRH record ID
128584
FDA classified
2014-08-19
Recall event number
68738
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.