During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly
dose audits/verification required to comply with the regulations could not be found. Sterilization
records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.
Source facts
Action
On 04/13/18, customer recall communication forms were sent via mail informing customers to quarantine effected devices, call the manufacturer to arrange device return at 800-633-7221, and report adverse events to the FDA's MedWatch.
Initiated
2018-04-13
Root cause
Process design
Terminated
2021-01-22
Distribution
U.S.: IL, UT, PA, NJ, WA, TX, NV, CA, NC, MS, KS, MO, MN, FL;
OUS: Australia and Sweden
Recall number
Z-2265-2018
Recall status
Terminated
CDRH record ID
164604
FDA record created
2018-06-20
Recall event number
80099