Open original FDA record ↗Recall record
- FDA identifier
- Z-2265-2019
- Company in FDA record
- Degania Silicone, Ltd.
- FDA record creation date
- Aug 12, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
Source facts
- Action
- Degania Silicone Ltd. notified accounts by letter dated June 28, 2019 describing the problem, health risk and action to take: Please forward this information to the immediate attention of the persons responsible for usage of the affected parts. Sub-recall must be initiated by your company to collect any distributed product from the suspected list.2. Please fill in the attached Recall Response form and indicate the quantity of the affected product in your possession. 3. Promptly return the completed Recall Response form to Degania Silicone by e-mail: kkayam@qco.net. 4. Disposition of the recalled product: - the preferred option for us is to scrap any recalled product at your facility. In this case we will require official Destruction Confirmation letter containing details of the scrapped product, signed by your company (the draft will be provided by request). - if the above option is not acceptable, we would ask to return any recalled product to our Rhode Island facility. Questions contact Degania Silicone: Degania Silicone Ltd. Degania Bet, 1513000, Israel Katia Kayam, Customer complaints Officer Tel: +972-529532059 E-mail: Kkayam@qco.net
- Initiated
- 2019-06-28
- Root cause
- Process control
- Terminated
- 2020-10-22
- Distribution
- Nationwide Distribution
- Recall number
- Z-2265-2019
- Recall status
- Terminated
- CDRH record ID
- 174959
- FDA record created
- 2019-08-12
- Recall event number
- 83287