Skip to content

Public FDA record

Z-2265-2025: Brand Name: T4 Total ELISA Product Name: T4 Total ELISA Model/Catalog Number: EIA-4568 Product Descriptio

Recall record Z-2265-2025 for DRG International, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2265-2025
Company in FDA record
DRG International, Inc.
FDA classification date
Aug 4, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

An incoming complaint reported that the expiration date on the kit box label and Certificate of Analysis (CoA) exceeded the shelf life of the standards and control solutions in the kit by 2 months (i.e., 2026-03-31 instead of 2026-01-31). The expiration dates on the bottle labels were correct (i.e., 2026-01-31).

Source facts

Action
On June 4, 2025 FIELD SAFETY NOTICE letters were sent to customers. Action to be taken by distributors and end-users: Please check your inventory for the respective lot. Please use the CoA provided with this FSN. If you require corrected outer kit box label for your documentation, please contact DRG. If you have transferred any of the affected products to another laboratory, please provide them with a copy of this letter. If you have any questions concerning this issue, please do not hesitate to contact us.
Initiated
2025-06-04
Root cause
Reprocessing Controls
Distribution
Worldwide - US Nationwide distribution in the states of CA, MA, PA and the countries of Germany, Hong Kong, Moldova.
Recall number
Z-2265-2025
Recall status
Open, Classified
CDRH record ID
214816
FDA classified
2025-08-04
Recall event number
97196

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗