Z-2266-2012: Toshiba CT systems, Asteion and Aquilion, that are equipped with a PC based computer, or to those older s

Recall record Z-2266-2012 for Toshiba American Medical Systems Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2266-2012
Company in FDA record
Toshiba American Medical Systems Inc
FDA classification date
Aug 24, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

The systems all have a similar table control board. It has been found that this board may cause the table position to be incorrect if the table cradle is moved manually. This is referred to as "Free" mode in the operation manuals. If this error does occur it is possible that the patient could be scanned at an incorrect position.

Source facts

Action
Toshiba sent a "URGENT: MEDICAL DEVICE CORRECTION" letter to all affected customers. The letter identified the product, the problem, and the action needed to be taken by the customer. Customers were instructed to please avoid moving the table in the "Free" or manual mode as much as possible. If it is necessary to move the table in this manner please verify the table position prior to initiating the scan sequence. If it is noted that the position is not correct, exit the eXam plan and start a new one beginning with the scanoscopy. Please share this information with all users of the equipment and your clinical or biomedical engineering departments. Please find the attached customer reply form. The purpose of this form is for FDA tracking. Please take a moment to complete the form and fax it to the toll free number on the form.". The contact information for this recall is:"(800) 421-1968 or contact your local Toshiba Representative at (800) 521-1968.".
Initiated
2008-02-27
Root cause
Process design
Terminated
2012-09-04
Distribution
USA ( Nationwide Distribution )
Recall number
Z-2266-2012
Recall status
Terminated
CDRH record ID
111873
FDA classified
2012-08-24
Recall event number
62851

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗