Skip to content

Public FDA record

Z-2266-2024: VITROS Chemistry Products FS Diluent Pack 3, Model/Catalog Number: 6801754. Product Description: VITROS C

Recall record Z-2266-2024 for Ortho-Clinical Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2266-2024
Company in FDA record
Ortho-Clinical Diagnostics, Inc.
FDA classification date
Jul 3, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

VITROS Chemistry Products FS Diluent Pack 3, Lot 01-1645 may cause results to be negatively biased when used to dilute patient samples tested with VITROS Chemistry Products CRP Slides.

Source facts

Action
An URGENT PRODUCT CORRECTION NOTIFICATION dated 5/31/24 was sent to customers. REQUIRED ACTIONS " Discontinue using, render unusable, and discard your remaining inventory of VITROS FS Diluent Pack 3, Lot 01-1645, including packs currently loaded on your VITROS System(s). " Complete and return the enclosed Confirmation of Receipt form no later than June 10, 2024. Upon receipt of your completed Confirmation of Receipt form, QuidelOrtho will credit or replace your discarded inventory. " Save this notification with your User Documentation or post this notification by each VITROS 5,1 FS/4600/5600/XT 7600 System until the issue has been resolved. " Please forward this notification if the affected product was distributed outside of your facility. " If your laboratory has experienced the issue with this product and you have not already done so, please report the occurrence to your local Global Services Organization. DISTRIBUTOR REQUIRED ACTIONS " Discard your remaining inventory of VITROS FS Diluent Pack 3, Lot 01-1645. QuidelOrtho will credit your account. " Send the enclosed customer letter and customer Confirmation of Receipt form to all customers who were shipped VITROS FS Diluent Pack 3, Lot 01-1645, from your facility. " Complete and return the enclosed distributor Confirmation of Receipt form no later than June 10, 2024. " Please forward this notification if the affected product was distributed outside of your facility. Contact Information We apologize for the inconvenience this may cause your facility. If you have further questions, please contact Global Services Organization at 1-800-421-3311.
Initiated
2024-05-21
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution in the states of FL, IL, NY, PA and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Columbia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Spain, Singapore, Sweden, The Netherlands, United Kingdom.
Recall number
Z-2266-2024
Recall status
Open, Classified
CDRH record ID
208223
FDA classified
2024-07-03
Recall event number
94770

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗