Open original FDA record ↗Recall record
- FDA identifier
- Z-2267-2024
- Company in FDA record
- Angiodynamics, Inc.
- FDA classification date
- Jul 3, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A single batch of ISO-100 IsoLoc Gas-Release Endorectal Balloons is mislabeled. The affected ISO-100 balloon catheter has a nominal fill volume of 100cc/ml but the inner pouch identifies the product as an ISO-60 (fill volume of 60cc/ml).
Source facts
- Action
- On May 21, 2024 URGENT VOLUNTARY MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: IMMEDIATELY: 1. Stop using the product subject to recall. 2. Remove any affected (recalled) product from your inventory (whether in labs, Central Supply, Shipping and Receiving or ANY other location). 3. Segregate this product in a secure location for return to AngioDynamics, Inc. 4. Forward a copy of this recall notification to all sites to which you have distributed affected product.
- Initiated
- 2024-05-21
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide distribution in the states of AZ, FL, IL, MO, MS, NJ, NM, NY, PA, TX, WI.
- Recall number
- Z-2267-2024
- Recall status
- Open, Classified
- CDRH record ID
- 208254
- FDA classified
- 2024-07-03
- Recall event number
- 94773