Skip to content

Public FDA record

Z-2267-2024: IsoLoc Prostate Immobilization Treatment Balloon Device Kit, Catalog/UPN: ISO-100

Recall record Z-2267-2024 for Angiodynamics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2267-2024
Company in FDA record
Angiodynamics, Inc.
FDA classification date
Jul 3, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A single batch of ISO-100 IsoLoc Gas-Release Endorectal Balloons is mislabeled. The affected ISO-100 balloon catheter has a nominal fill volume of 100cc/ml but the inner pouch identifies the product as an ISO-60 (fill volume of 60cc/ml).

Source facts

Action
On May 21, 2024 URGENT VOLUNTARY MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: IMMEDIATELY: 1. Stop using the product subject to recall. 2. Remove any affected (recalled) product from your inventory (whether in labs, Central Supply, Shipping and Receiving or ANY other location). 3. Segregate this product in a secure location for return to AngioDynamics, Inc. 4. Forward a copy of this recall notification to all sites to which you have distributed affected product.
Initiated
2024-05-21
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of AZ, FL, IL, MO, MS, NJ, NM, NY, PA, TX, WI.
Recall number
Z-2267-2024
Recall status
Open, Classified
CDRH record ID
208254
FDA classified
2024-07-03
Recall event number
94773

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.