Z-2267-2025: DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71457

Recall record Z-2267-2025 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2267-2025
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Aug 5, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.

Source facts

Action
An URGENT MEDICAL DEVICE RECALL notification dated 6/25/25 was sent to customers. ACTION: If your laboratory encounters any of the events described in this letter: " Perform a system initialization to resume normal operation. o If repeated initialization attempts fail, contact Customer Technical Support. " Repeat testing for any cancelled tests. " No action is required if your laboratory has not observed the events listed above ESOLUTION: "A Beckman Coulter representative will schedule the modification installation when it is available. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: " Electronically, if you received this communication via email. " Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact our Customer Support: " From our website: http://www.beckmancoulter.com
Initiated
2025-06-25
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide.
Recall number
Z-2267-2025
Recall status
Open, Classified
CDRH record ID
214694
FDA classified
2025-08-05
Recall event number
97162

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗