Open original FDA record ↗Recall record
- FDA identifier
- Z-2269-2010
- Company in FDA record
- Essential Medical Supply, Inc.
- FDA classification date
- Aug 25, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Essential Medical Supply, Inc. is recalling W1801B and W1801R Endurance HD Walkers due to recent reports and tests that show the front caster may detach from the frame. The potential failures include a collapse of the plastic fork or shearing of the stud attaching the caster to the frame.
Source facts
- Action
- All Consignees were mailed an Urgent: Product Recall W1801B and W1801R Endurance HD Walkers notification on December 30, 2009. All notifications were sent first class mail in a 9" X 6" envelope. Response cards were included. The letter identified the affected product and the reason for recall. Customers were asked to examine their inventory for affected product and to discontinue further use or distribution of the product. If the product was further distributed, customers should contact their customers, advise them of the recall and have them return the units to them. Customers should follow the instructions provided for return and replacement of units.
- Initiated
- 2009-12-21
- Root cause
- Other
- Terminated
- 2012-02-21
- Distribution
- Nationwide Distribution
- Recall number
- Z-2269-2010
- Recall status
- Terminated
- CDRH record ID
- 89229
- FDA classified
- 2010-08-25
- Recall event number
- 54810