Issuance of validated, new reprocessing procedures.
This corrective action is being undertaken as a result of ongoing manufacturer and FDA activities relating to reported patient infections associated with duodenoscopes. The new cleaning procedure requires additional recess flushing and forceps elevator raising/lowering steps during precleaning, manual cleaning and manual disinfection.
Source facts
Action
Firm is extending recall to the user level
Letters will be sent on May, 11, 2018 via overnight mail to customers
Follow-up phone calls to consignees to offer training on new reprocessing instructions
No action can be taken with hospitals that are no longer in business
Initiated
2018-05-11
Root cause
Under Investigation by firm
Terminated
2020-05-05
Distribution
Nationwide
Recall number
Z-2269-2018
Recall status
Terminated
CDRH record ID
164185
FDA record created
2018-06-21
Recall event number
79998