Potential for four safety-related software issues impacting systems utilizing syngo.CT software version VB10A which may result in a scan abort.
Source facts
Action
Siemens has initiated a Customer Safety Advisory Notice (CSAN) to all affected customers via CT035/19/S. Customers with valid email addresses were emailed a copy of the CSAN on 07/03/2019 via Adobe Mega Sign. The CSAN were distributed to the remaining customers without a valid email address via certified mail on 07/12/2019. Additionally, a solution to eliminate the root cause of this problem (application of software update VB20) will be distributed to all affected customers free of charge. Following the corrective action, the cause is eliminated and a recurrence of the fault is prevented.
Initiated
2019-07-03
Root cause
Software design
Terminated
2024-05-10
Distribution
US Nationwide distribution.
Recall number
Z-2269-2019
Recall status
Terminated
CDRH record ID
174998
FDA record created
2019-08-14
Recall event number
83377