Skip to content

Public FDA record

Z-2270-2023: Over-Head Ceiling Tube Support, CH-200, CH-200M, Used with RadSpeed and RadSpeed PRO, Sonial Vision safir

Recall record Z-2270-2023 for SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2270-2023
Company in FDA record
SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION
FDA classification date
Jul 31, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Eight fixing bolts at the base of the X-ray ceiling tube column assembly may be loose or missing so service is required to install a cable, bracket, and label behind the bracket, indicating when service is required, to eliminate the risk of the device falling.

Source facts

Action
On 4/10/23, recall notices were emailed to customers who were asked to do the following: 1)The Tube Column Assembly is secured by eight fixing bolts located at the base of the assembly. If while using the system, you observe loose, or missing bolts then please discontinue use of the system and call your authorized Service Provider. 2) Schedule a time with their Distributer to have their system updated. 3) Post the notice within their facility so that all users are aware of how to avoid this issue until the corrective actions has been completed. 4) Complete and return the Customer Acknowledgement and Receipt of Letter form to drodriguez@shimadzu-usa.com The National Technical Support phone number was provided: 844-487-2767
Initiated
2023-04-10
Root cause
Device Design
Distribution
U.S. Nationwide distribution.
Recall number
Z-2270-2023
Recall status
Open, Classified
CDRH record ID
200687
FDA classified
2023-07-31
Recall event number
92391

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗