Open original FDA record ↗Recall record
- FDA identifier
- Z-2271-2015
- Company in FDA record
- Draeger Medical, Inc.
- FDA classification date
- Jul 27, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
the Fabius MRI or parts of the system were attracted by the magnetic field of the MRI. If the Fabius MRI is positioned too close to the MRI, its parts can be loosened as a result of the magnetic force of the MRI.
Source facts
- Action
- Draeger Medical sent an Urgent Medical Device Recall Letter, dated June 2015, and Instructions for Use Supplement (one supplement for each Fabius MRI Anesthesia Machine at the facility) to end users. The letter identified the affected product, problem and actions to be taken. For questions regarding the recall letter contact Michael Kelhart at 1-800-543-5047 (press 1 at the prompt, then 2, then 32349). For questions regarding the operation and/or servicing of the Drger Fabius MRI anesthesia machine contact DrgerService Technical Support at 1-800-543-5047 (press 4 at the prompt).
- Initiated
- 2015-06-22
- Root cause
- Labeling False and Misleading
- Terminated
- 2017-05-05
- Distribution
- US Nationwide Distribution
- Recall number
- Z-2271-2015
- Recall status
- Terminated
- CDRH record ID
- 138314
- FDA classified
- 2015-07-27
- Recall event number
- 71592