Open original FDA record ↗Recall record
- FDA identifier
- Z-2273-2016
- Company in FDA record
- Ortho-Clinical Diagnostics
- FDA record creation date
- Jul 23, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Ortho Clinical Diagnostics (OCD) confirmed a potential for biased calcium (Ca) results to be generated when using VITROS Ca Slides, GEN29, and the VITROS 250/350/950/5,1 FS, 4600 Chemistry System or the VITROS 5600 Integrated System. An increase in negatively biased calcium results complaints prompted the investigation that identified this product issue.
Source facts
- Action
- Ortho Clinical Diagnostics sent an URGENT PRODUCT CORRECTION NOTIFICATION letter dated May 18, 2016 to affected customers. . The letter identified the affected product, problem and actions to be taken. The letter instructed to discontinue use of the affected lot(s) and discard all remaining inventory. Customers were asked to complete and return the Confirmation of Receipt form. For questions contact Customer Technical Services at 1-800-421-3311.
- Initiated
- 2016-05-18
- Root cause
- Nonconforming Material/Component
- Terminated
- 2018-06-08
- Distribution
- Worldwide Distribution - US Nationwide in the countries of Puerto Rico and Foreign distribution to the following countries: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, Poland and Venezuela.
- Recall number
- Z-2273-2016
- Recall status
- Terminated
- CDRH record ID
- 147262
- FDA record created
- 2016-07-23
- Recall event number
- 74355