Sterility not assured. The stopcock handle position was inadvertently changed from an open to a closed position, which prevents sterilization of the closed portion of the device. Sterility of the affected product cannot be guaranteed.
Source facts
Action
June 7, 2010 Notification letter to customer instructed them to return the product to Arrow. Arrow will destroy the returned product.
Initiated
2010-06-07
Root cause
Process change control
Terminated
2012-09-29
Distribution
All affected product was distributed to one foreign consignee: Arrow Japan Ltd.
Recall number
Z-2274-2010
Recall status
Terminated
CDRH record ID
92259
FDA classified
2010-08-25
Recall event number
55933