Z-2274-2025: DASH SARS-CoV-2 & Flu A/B Test Model/Catalog Number: SG-0006 combination COVID-19 and Flu test. Eac

Recall record Z-2274-2025 for Nuclein LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2274-2025
Company in FDA record
Nuclein LLC
FDA classification date
Aug 6, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to manufacturing error, assay test may result in missed diagnosis or delay of treatment. (Potential false negative result)

Source facts

Action
On June 25, 2025, Nuclein, LLC issued a Urgent: Medical Device Recall notification to affected customers. On July 8, 2025 an updated notification was issued via E.Mail. a. Review your inventory of DASH SARS-CoV-2 & Flu A/B Tests to confirm the quantity of Lot BB02 or BB03 that you have remaining. b. Suspend the use and quarantine the affected lot of DASH SARS-CoV-2 & Flu A/B Tests; Nuclein will replace any unused affected product. c. Review all results of the product lots detailed in Section 6, below, to identify potentially erroneous diagnosis, except in those cases where alternative confirmation was obtained. d. Forward this information to all individuals and departments within your organization using the product information detailed in Section 6. If you are not the end user, forward this notice to the product end user. e. Complete the Acknowledgement and Receipt Form below and either email the completed form to customersupport@nuclein.com or mail to the address identified in the Acknowledgement and Receipt Form. f. Return the affected tests that you have remaining; Nuclein Customer Support will follow up with you on the details of returning or destroying the affected product.
Initiated
2025-06-25
Root cause
Process control
Distribution
US distribution to: Florida, New York, Texas, and West Virginia
Recall number
Z-2274-2025
Recall status
Open, Classified
CDRH record ID
214936
FDA classified
2025-08-06
Recall event number
97228

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗