Open original FDA record ↗Recall record
- FDA identifier
- Z-2275-2010
- Company in FDA record
- GE Healthcare, LLC
- FDA classification date
- Aug 25, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
GE Healthcare has become aware of four potential safety issues associated with the CARESCAPE" Monitor B850: 1. Loss of user input when using certain displays or the USB remote control. 2. Potential for a delay in treatment or missed alarm when the alarm light is used as a primary alarm source and audible alarms are turned down or off. 3. The CARESCAPE Monitor B850 will not use user entered hemo
Source facts
- Action
- Consignees were sent on 7/16/10, a GE "Urgent Medical Device Correction" letter dated July 12, 2010. The letter was addressed to Healthcare Administrator/Risk Manger, Chief of Nursing, Director of Biomedical Engineering. They stated the Safety Issues, Affected Product Details, Product Correction, Safety Instructions and Contact Information. If you have any questions, please contact Technical Support at 1-800-558-7044.
- Initiated
- 2010-07-16
- Root cause
- Software design
- Terminated
- 2013-09-18
- Distribution
- CA, CO, FL, GA, KY, LA, MA, MI, NE, NJ, NY, NC, OH, OR, PA, TX VA, and WI, Foreign: UNITED KINGDOM, TURKEY, SWITZERLAND, SPAIN , SLOVENIA, SINGAPORE, SAUDI ARABIA, PORTUGAL NETHERLAND, ITALY, GERMANY, FRANCE,FINLAND, DENMARK, CZECH REPUBLIC, BELGIUM, AUSTRIA, and AUSTRALIA.
- Recall number
- Z-2275-2010
- Recall status
- Terminated
- CDRH record ID
- 92731
- FDA classified
- 2010-08-25
- Recall event number
- 56141