Open original FDA record ↗Recall record
- FDA identifier
- Z-2276-2013
- Company in FDA record
- Elekta, Inc.
- FDA classification date
- Sep 25, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Upgrade installation script is designed to overwrite the same pouch files in the Treatment Couch Library with the original default values.
Source facts
- Action
- Monaco 3.3 upgrades have been temporarily halted and Important Safety Notice (LINMON0007) has been issued to effected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to ensure that couch files are checked for releases 3.20.00 and 3.30.00 to verify that the correct relative electron density values are being used in the dose calculation. If customers have edited the sample couch files, and previously upgraded to Monaco 3.20.00 or 3.30.00, their couch files have already been overwritten. Open the couch files, make the necessary updates to the couch parameters, and use the "Save As Treatment Couch" option to save the modified sample couch files to a new, unique name. Please note that the couch files can also be edited in Focal and then the patients opened in Monaco. Comparable Focal releases would be Focal4.64.00(Monaco 3.10.00), Focal4.70.00(Monaco 3.20.00) and Focal4.80.00 (Monaco 3.30.00). For questions regarding this recall call 404-993-5886.
- Initiated
- 2013-07-02
- Root cause
- Device Design
- Terminated
- 2014-05-08
- Distribution
- Nationwide Distribution including AK, CA, CO, CT, FL, IL, IN, MI, MN, MO, NC, NJ, NY, OH, OK, OR, PA, SC, TN, TX, WA and Puerto Rico.
- Recall number
- Z-2276-2013
- Recall status
- Terminated
- CDRH record ID
- 121167
- FDA classified
- 2013-09-25
- Recall event number
- 65996