Open original FDA record ↗Recall record
- FDA identifier
- Z-2276-2019
- Company in FDA record
- NeuroLogica Corporation
- FDA record creation date
- Aug 14, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a potential for probes overheating when decreasing the Doppler SV (Sample Volume) size value set by the user in the Doppler Only mode.
Source facts
- Action
- On July 25, 2019, Neurological Corporation a subsidiary of Samsung Electronics issued Urgent Field Safety Notices to customers. Customers were advised to the following actions: 1) If the user wants to decrease the Dopper SV (Sample Volume) size set by himself/herself in the Doppler Only Mode, the user should press "set" button to update the PW Doppler mode. 2) A copy of the Field Safety Notice should be passed on to all who need to be aware within your organization. 3) Contact your local Samsung representative if you have questions. 4) Return the completed Customer Reply Form to Tom Leinart tleinart@neurologica.com.
- Initiated
- 2019-07-25
- Root cause
- Software design
- Terminated
- 2020-09-11
- Distribution
- US Nationwide
- Recall number
- Z-2276-2019
- Recall status
- Terminated
- CDRH record ID
- 175193
- FDA record created
- 2019-08-14
- Recall event number
- 83429