Z-2276-2021: Alinity s System REF 06P16-01. An automated immunoassay analyzer for In Vitro diagnostic use in blood d

Recall record Z-2276-2021 for Abbott Laboratories, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2276-2021
Company in FDA record
Abbott Laboratories, Inc
FDA record creation date
Aug 12, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A design defect (hardware and software) allows liquid waste pressure to build up and potentially spray users/operators.

Source facts

Action
On 07/12/2021, the firm started hand-delivering an "Product Correction Urgent-Immediate Action Required" Notification to customers informing them that the Recalling Firm has identified an issue with pressure build up in the external waste tubing of the liquid waste assembly system if obstructed leading to a potential for biohazardous exposure and/or operator injury (contact with liquid waste). Customers are instructed to follow the below instruction until the Recalling Firm can upgrade the analyzers with a future system software version and a hardware component: 1. As stated in Section 8, Hazards, follow all product labeling and safety instructions in the laboratory. Always wear personal protective equipment (PPE) when operating the analyzer system, other laboratory instrumentation and equipment, and when using chemicals. 2. Adhere to the instructions listed in the Operations Manual, Section 10, to resolve message codes pertaining to waste overflow and contact Customer Service to resolve issues related to hardware failure. 3. Use caution when handling the Alinity s System external waste tubing. Avoid any configuration that could impact the free flow of the external waste and cause pressure buildup within the external waste tubing. A future software update and associated hardware upgrade will be implemented to enable external liquid waste tubing pressure monitoring and the display of the expected message code 5101 Liquid waste high pressure detected. 4. If customers have forwarded the product listed above to other laboratories, please inform them of this Product Correction and provide to them a copy of this letter. 5. Complete and return the Customer Reply form. 6. Retain this letter for their laboratory records. If customers or any of the health care providers have questions regarding this information, U.S. Customers contact Customer Service at 1-877-4ABBOTT (available 24 hours a day, 7 days a week). Customers outside the U.S., please contact your local area Cust
Initiated
2021-07-12
Root cause
Software design
Distribution
Worldwide distribution. US nationwide, Australia, Bahrain, Belgium, Brazil, China, Colombia, Croatia, Egypt, Germany, Greece, Hong Kong, Hungary, Iran, Ireland, Israel, Italy, Jordan, Kuwait, Mexico, Netherlands, Pakistan, Poland, Portugal, Saudi Arabia, Serbia, Slovenia, South Africa, Spain, Tanzania, United Arab Emirates, United Kingdom, Ukraine and Zimbabwe.
Recall number
Z-2276-2021
Recall status
Open, Classified
CDRH record ID
188388
FDA record created
2021-08-12
Recall event number
88324

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗