Open original FDA record ↗Recall record
- FDA identifier
- Z-2277-2025
- Company in FDA record
- Stryker Corporation
- FDA classification date
- Aug 7, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Tourniquet cuff flange may become detached from the bladder. If leak occurs during the procedure, there may be risk of hemorrhage, resulting in blood loss greater than expected. Other serious risks include cerebral hypotension and cardiovascular collapse and arrest, risking stroke or death. Specific to IVRA, patient toxicity may also occur due to rapid systemic absorption of anesthetics.
Source facts
- Action
- On July 2, 2025, the firm notified customers via Urgent Medical Device Recall letters. Customers were instructed to locate and quarantine any affected products at their facility and return an enclosed business reply form to the firm. Stryker will provide a shipping label to return the affected product, and credit will be issued to affected customers upon receipt of the recalled product.
- Initiated
- 2025-07-02
- Root cause
- Equipment maintenance
- Distribution
- Domestic US distribution nationwide. International distribution to Australia Canada Hong Kong India Malaysia Netherlands New Zealand Singapore Sweden Thailand United Kingdom
- Recall number
- Z-2277-2025
- Recall status
- Open, Classified
- CDRH record ID
- 214628
- FDA classified
- 2025-08-07
- Recall event number
- 97133