Z-2278-2021: Edwards PediaSat Oximetry Catheter Kit, REF: XT3515KTP, Exterior Diameter 5.5F (1.83 mm), Usable Length 1

Recall record Z-2278-2021 for Edwards Lifesciences, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2278-2021
Company in FDA record
Edwards Lifesciences, LLC
FDA record creation date
Aug 12, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for internal leaks within catheters.

Source facts

Action
On 7/9/2021 Edwards issued a "URGENT MEDICAL DEVICE PROUDCT CORRECTION" via FedEx to all affected consignees. In addition to informing consignees about the recall the firm ask consignees take the following actions: 1. The problem may be identified during preparation of the catheter for insertion per the IFU, paying specific attention to the connection between the green hub and the tubing for any signs of leakage. Additionally, the problem may be identified if when flushing the lumens of the catheter while administering a medication, a transient change of vital signs is observed. In either of these situations, remove the catheter over a guidewire maintaining venous access and then place a new product over the venous wire. Once properly placed, remove the guidewire, and secure the catheter. New venous access is not required. 2. Please follow the instructions included in the enclosed acknowledgement form and listed below. Return the acknowledgement form to US.FCA@edwards.com. 3. There is no need to return product at this time. 4. Distribute this notice within your organization or to any organization where the potentially affected devices have been transferred. 5. E-mail the completed form to Edwards Customer Service at US.FCA@edwards.com within 5 days from receipt of this notification. Product can be used as indicated above. If you have any questions, you may call Customer Service at 1-800-424-3278 option 1, from 5AM 4:30PM PST. 6. Advice on action to be taken by Distributor: 7. Please review this letter and complete the acknowledgement form. Return the acknowledgement form to US.FCA@edwards.com. Please forward this customer communication to any of your customers who have purchased the impacted Edwards PediaSat product and ask them to return the enclosed acknowledgement form. 8. Alternatively, you can provide Edwards with a list of your customers who have purchased the impacted product and Edwards will communicate directly with your customers to facilit
Initiated
2021-07-09
Root cause
Process control
Terminated
2024-01-29
Distribution
US: CA, TX, FL, AZ, PA, IN, NY, NJ, MI, WA, MD, OH, FL, OUS: Japan, Mexico, Brazil, Colombia, Taiwan, India, Poland, Germany, United Arab Emirates, Israel, Switzerland, Bahrain, Iman, Tunisia, Austria, Belgium, France, Netherlands, Italy, Spain, Canary Islands, Portugal, Greece, Costa Rica, Canada
Recall number
Z-2278-2021
Recall status
Terminated
CDRH record ID
188260
FDA record created
2021-08-12
Recall event number
88270

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗