Skip to content

Public FDA record

Z-2278-2025: FARAWAVE 1.0 Pulsed Field Ablation Catheter, Product ID M004PF41M401

Recall record Z-2278-2025 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2278-2025
Company in FDA record
Boston Scientific Corporation
FDA classification date
Aug 8, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The potentially impacted units were manufactured using specific equipment that may have caused cracks in the electrode bands on the catheter's distal end.

Source facts

Action
An Urgent Medical Device Removal - Immediate Action Required notification letter dated June 2025 was sent to customers. Actions 1. Do NOT use affected product (Attachment 1), remove affected devices from your facility s inventory, segregate the units in a secure place until they can be returned to Boston Scientific. 2. Immediately post this information in a visible location near the affected products to ensure this information is readily accessible to all handlers and users of the device. 3. Forward this notice to any healthcare professional from your organization for awareness and to any organization where affected devices have been transferred. 4. Complete and return the enclosed Reply Verification Tracking Form per the enclosed instructions on page five. 5. Return affected product. Email: BSCFieldActionCenter@bsci.com Reach out to your local Boston Scientific representative with any questions.
Initiated
2025-06-25
Root cause
Nonconforming Material/Component
Distribution
US Nationwide.
Recall number
Z-2278-2025
Recall status
Open, Classified
CDRH record ID
214843
FDA classified
2025-08-08
Recall event number
97207

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗