Z-2279-2023: Field Upgrade Kits: P/N 5451208 RIU, 9 INCH, TOSHIBA; P/N 5485304 SERVICE PART, IMAGE INTENSIFIER, 9 INCH

Recall record Z-2279-2023 for GE OEC Medical Systems, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2279-2023
Company in FDA record
GE OEC Medical Systems, Inc
FDA classification date
Jul 31, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.

Source facts

Action
On 5/15/23 correction notices were mailed to: Director/Manager of Radiology, Hospital Administrator, Head of Radiology Department, PACS Administrator, Director of IT Department, Head, Biomedical Engineering, who were asked to do the following: You can continue to use your device. When transporting the device, ensure it is always in the Anterior Posterior (AP) position, as instructed in the systems Operator Manual. Use caution when moving system to avoid a large impact force. In addition, customers were asked to complete and return the acknowledgement response form to fieldactionssurgery@ge.com Customers with questions or concerns can contact the firm at 1-800-437-1171
Initiated
2023-05-15
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Algeria, Argentina, Australia, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Burkina Faso, Canada, Chile, Colombia, Czechia, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, France, Georgia, Germany, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Kazakhstan, KOREA, Malaysia, Mauritius, Mexico, Mongolia, Morocco, Nepal, New Caledonia, New Zealand, Nigeria, Paraguay, Peru, Poland, Portugal, Romania, Russia, Singapore, Spain, Switzerland, Taiwan, Tanzania, Thailand, Trinidad and Tobago, United Arab Emirates, United Kingdom, Uruguay, Vietnam.
Recall number
Z-2279-2023
Recall status
Open, Classified
CDRH record ID
201874
FDA classified
2023-07-31
Recall event number
92504

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗